D.E. = | Average weight recovered (mg) ——————————————— weight added (mg) | |
(VI) The desorption efficiency is dependent on the amount of analyte collected on the charcoal. Plot the desorption efficiency versus weight of analyte found. This curve is used in (j)(iv) of this subsection to correct for adsorption losses.
(i) Calibration and standards. It is convenient to express concentration of standards in terms of mg/1.0 ml methanol, because samples are desorbed in this amount of methanol. The density of the analyte is used to convert mg into microliters for easy measurement with a microliter syringe. A series of standards, varying in concentration over the range of interest, is prepared and analyzed under the same GC conditions and during the same time period as the unknown samples. Curves are established by plotting concentration in mg/1.0 ml versus peak area.
Note: | Since no internal standard is used in the method, standard solutions must be analyzed at the same time that the sample analysis is done. This will minimize the effect of known day-to-day variations and variations during the same day of the FID response. |
(j) Calculations.
(i) Read the weight, in mg, corresponding to each peak area from the standard curve. No volume corrections are needed, because the standard curve is based on mg/1.0 ml methanol and the volume of sample injected is identical to the volume of the standards injected.
(ii) Corrections for the bank must be made for each sample.
Where: |
mg sample = mg found in front section of sample tube. |
mg sample = mg found in front section of blank tube. |
Note: | A similar procedure is followed for the backup sections. |
(iii) Add the weights found in the front and backup sections to get the total weight in the sample.
(iv) Read the desorption efficiency from the curve (reference (h)(v)(B) of this subsection) for the amount found in the front section. Divide the total weight by this desorption efficiency to obtain the corrected mg/sample.
Corrected mg/sample = | Total weight ———— D.E. | |
(v) The concentration of the analyte in the air sampled can be expressed in mg/cu m.
mg/cu m = Corrected mg (see (j)(iv)) x | 1,000 (liter/cu m) —————————————— air volume sampled (liter) |
(vi) Another method of expressing concentration is ppm.
ppm = mg/cu m × 24.45/M.W. × 760/P × T + 273/298 |
Where: |
P = Pressure (mm Hg) of air sampled. |
T = Temperature (°C) of air sampled. |
24.45 = Molar volume (liter/mole) at 25°C and 760 mm Hg. |
M.W. = Molecular weight (g/mole) of analyte. |
760 = Standard pressure (mm Hg). |
298 = Standard temperature (°K). |
(k) References.
(i) White, L. D. et al., "A Convenient Optimized Method for the Analysis of Selected Solvent Vapors in the Industrial Atmosphere," Amer. Ind. Hyg. Assoc. J., 31:225 (1970).
(ii) Documentation of NIOSH Validation Tests, NIOSH Contract No. CDC-99-74-45.
(iii) Final Report, NIOSH Contract HSM-99-71-31, "Personal Sampler Pump for Charcoal Tubes," September 15, 1972.
(7) NIOSH Modification of NIOSH Method S-156. The NIOSH recommended method for low levels for acrylonitrile is a modification of method S-156. It differs in the following respects:
(a) Samples are desorbed using 1 ml of 1 percent acetone in CS2 rather than methanol.
(b) The analytical column and conditions are:
(i) Column: 20 percent SP-1000 on 80/100 Supelcoport 10 feet × 1/8 inch S.S.
(ii) Conditions:
Injector temperature: 200°C.
Detector temperature: 100°C.
Column temperature: 85°C.
Helium flow: 25 ml/min.
Air flow: 450 ml/min.
Hydrogen flow: 55 ml/min.
(c) A 2 µl injection of the desorbed analyte is used.
(d) A sampling rate of 100 ml/min is recommended.
(8) OSHA Laboratory Modification of NIOSH Method S-156.
(a) Analyte: Acrylonitrile.
(b) Matrix: Air.
(c) Procedure: Adsorption on charcoal, desorption with methanol, GC.
(d) Principle of the method (subsection (1)(a) of this section).
(i) A known volume of air is drawn through a charcoal tube to trap the organic vapors present.
(ii) The charcoal in the tube is transferred to a small, stoppered sample vial, and the analyte is desorbed with methanol.
(iii) An aliquot of the desorbed sample is injected into a gas chromatograph.
(iv) The area of the resulting peak is determined and compared with areas obtained for standards.
(e) Advantages and disadvantages of the method.
(i) The sampling device is small, portable, and involves no liquids. Interferences are minimal, and most of those which do occur can be eliminated by altering chromatographic conditions. The tubes are analyzed by means of a quick, instrumental method.
(ii) This method may not be adequate for the simultaneous analysis of two or more substances.
(iii) The amount of sample which can be taken is limited by the number of milligrams that the tube will hold before overloading. When the sample value obtained for the backup section of the charcoal tube exceeds 25 percent of that found on the front section, the possibility of sample loss exists.
(iv) The precision of the method is limited by the reproducibility of the pressure drop across the tubes. This drop will affect the flow rate and cause the volume to be imprecise, because the pump is usually calibrated for one tube only.
(f) Apparatus.
(i) A calibrated personal sampling pump whose flow can be determined within ±5 percent at the recommended flow rate.
(ii) Charcoal tubes: Glass tube with both ends flame sealed, 7 cm long with a 6 mm O.D. and a 4 mm I.D., containing 2 sections of 20/40 mesh activated charcoal separated by a 2 mm portion of urethane foam. The activated charcoal is prepared from coconut shells and is fired at 600°C prior to packing. The absorbing section contains 100 mg of charcoal, the back-up section 50 mg. A 3 mm portion of urethane foam is placed between the outlet end of the tube and the back-up section. A plug of silicated glass wool is placed in front of the adsorbing section. The pressure drop across the tube must be less than one inch of mercury at a flow rate of 1 liter per minute.
(iii) Gas chromatograph equipped with a nitrogen phosphorus detector.
(iv) Column (10 ft × 1/8 in stainless steel) packed with 100/120 Supelcoport coated with 10 percent SP 1000.
(v) An electronic integrator or some other suitable method for measuring peak area.
(vi) Two-milliliter sample vials with Teflon-lined caps.
(vii) Microliter syringes: 10 microliter, and other convenient sizes for making standards.
(viii) Pipets: 1.0 ml delivery pipets.
(ix) Volumetric flasks: Convenient sizes for making standard solutions.
(g) Reagents.
(i) Chromatographic quality methanol.
(ii) Acrylonitrile, reagent grade.
(iii) Filtered compressed air.
(iv) Purified hydrogen.
(v) Purified helium.
(h) Procedure.
(i) Cleaning of equipment. All glassware used for the laboratory analysis should be properly cleaned and free of organics which could interfere in the analysis.
(ii) Calibration of personal pumps. Each pump must be calibrated with a representative charcoal tube in the line.
(iii) Collection and shipping of samples.
(A) Immediately before sampling, break the ends of the tube to provide an opening at least one-half the internal diameter of the tube (2 mm).
(B) The smaller section of the charcoal is used as the backup and should be placed nearest the sampling pump.
(C) The charcoal should be placed in a vertical position during sampling to minimize channeling through the charcoal.
(D) Air being sampled should not be passed through any hose or tubing before entering the charcoal tube.
(E) A sample size of 20 liters is recommended. Sample at a flow rate of approximately 0.2 liters per minute. The flow rate should be known with an accuracy of at least ±5 percent.
(F) The temperature and pressure of the atmosphere being sampled should be recorded.
(G) The charcoal tubes should be capped with the supplied plastic caps immediately after sampling. Rubber caps should not be used.
(H) Submit at least one blank tube (a charcoal tube subjected to the same handling procedures, without having any air drawn through it) with each set of samples.
(I) Take necessary shipping and packing precautions to minimize breakage of samples.
(iv) Analysis of samples.
(A) Preparation of samples. In preparation for analysis, each charcoal tube is scored with a file in front of the first section of charcoal and broken open. The glass wool is removed and discarded. The charcoal in the first (larger) section is transferred to a 2 ml vial. The separating section of foam is removed and discarded; the section is transferred to another capped vial. These two sections are analyzed separately.
(B) Desorption of samples. Prior to analysis, 1.0 ml of methanol is pipetted into each sample container. Desorption should be done for 30 minutes in an ultrasonic bath. The sample vials are recapped as soon as the solvent is added.
(C) GC conditions. The typical operating conditions for the gas chromatograph are:
(I) 30 ml/min (60 psig) helium carrier gas flow.
(II) 3.0 ml/min (30 psig) hydrogen gas flow to detector.
(III) 50 ml/min (60 psig) air flow to detector.
(IV) 200°C injector temperature.
(V) 200°C dejector temperature.
(VI) 100°C column temperature.
(D) Injection. Solvent flush technique or equivalent.
(E) Measurement of area. The area of the sample peak is measured by an electronic integator or some other suitable form of area measurement, and preliminary results are read from a standard curve prepared as discussed below.
(v) Determination of desorption efficiency.
(A) Importance of determination. The desorption efficiency of a particular compound can vary from one laboratory to another and also from one batch of charcoal to another. Thus, it is necessary to determine, at least once, the percentage of the specific compound that is removed in the desorption process, provided the same batch of charcoal is used.
(B) Procedure for determining desorption efficiency. The reference portion of the charcoal tube is removed. To the remaining portion, amounts representing 0.5X, 1X, and 2X (X represents TLV) based on a 20 l air sample are injected onto several tubes at each level. Dilutions of acrylonitrile with methanol are made to allow injection of measurable quantities. These tubes are then allowed to equilibrate at least overnight. Following equilibration they are analyzed following the same procedure as the samples. A curve of the desorption efficiency (amt recovered/amt added) is plotted versus amount of analyte found. This curve is used to correct for adsorption losses.
(i) Calibration and standards. A series of standards, varying in concentration over the range of interest, is prepared and analyzed under the same GC conditions and during the same time period as the unknown samples. Curves are prepared by plotting concentration versus peak area.
Note: | Since no internal standard is used in the method, standard solutions must be analyzed at the same time that the sample analysis is done. This will minimize the effect of known day-to-day variations and variations during the same day of the NPD response. Multiple injections are necessary. |
(j) Calculations. Read the weight, corresponding to each peak area from the standard curve, correct for the blank, correct for the desorption efficiency, and make necessary air volume corrections.
(k) Reference. NIOSH Method S-156.
[Statutory Authority: RCW
49.17.010,
49.17.040,
49.17.050, and
49.17.060. WSR 19-01-094, § 296-62-07340, filed 12/18/18, effective 1/18/19. Statutory Authority: Chapter
49.17 RCW. WSR 88-11-021 (Order 88-04), § 296-62-07340, filed 5/11/88.]