ETO, mg/m3 | = | AXB | |
C |
where: |
A | = | µg/mL |
B | = | desorption volume in milliliters |
C | = | air volume in liters. |
(E) To convert mg/m3 to parts per million (ppm) the following relationship is used:
ETO, ppm | = | mg/m3 x 24.45 | |
44.05 |
where: |
mg/m3 | = | results from 3.7.4 |
24.45 | = | molar volume at 25°C and 760 mm Hg |
44.05 | = | air volume in liters. |
(viii) Safety precaution
(A) Ethylene oxide and benzene are potential carcinogens and care must be exercised when working with these compounds.
(B) All work done with the solvents (preparation of standards, desorption of samples, etc.) should be done in a hood.
(C) Avoid any skin contact with all of the solvents.
(D) Wear safety glasses at all times.
(E) Avoid skin contact with HBr because it is highly toxic and a strong irritant to eyes and skin.
(i) Backup data.
(i) Detection limit data.
The detection limit was determined by injecting 0.8 µL of a 0.015 µg/mL standard of ethylene oxide into 1% CS2 in benzene. The detection limit of the analytical procedure is taken to be 1.20 x 10-5 µg per injection. This is equivalent to 8.3 ppb (0.015 mg/m3) for the recommended air volume.
(ii) Desorption efficiency. Ethylene oxide was spiked into charcoal tubes and the following recovery data was obtained:
Amount spiked (µg) | Amount recovered (µg) | Percent recovery |
4.5 | 4.32 | 96.0 |
3.0 | 2.61 | 87.0 |
2.25 | 2.025 | 90.0 |
1.5 | 1.365 | 91.0 |
1.5 | 1.38 | 92.0 |
.75 | 6525 | 87.0 |
.375 | .315 | 84.0 |
.375 | .312 | 83.2 |
.1875 | .151 | 80.5 |
.094 | .070 | 74.5 |
Note: | At lower amounts the recovery appears to be nonlinear. |
(iii) Sensitivity data. The following data was used to determine the calibration curve:
Injection | 0.5 x .75 µg/mL | 1 x 1.5 µg/mL | 2 x 3.0 µg/mL |
1 . . . . . . . . . . | 30904 | 59567 | 111778 |
2 . . . . . . . . . . | 30987 | 62914 | 106016 |
3 . . . . . . . . . . | 32555 | 58578 | 106122 |
4 . . . . . . . . . . | 32242 | 57173 | 109716 |
X . . . . . . . . . . | 31672 | 59558 | 108408 |
Slope = 34.105.
(iv) Recovery. The recovery was determined by spiking ethylene oxide onto lot 120 charcoal tubes and desorbing with 1% CS2 in Benzene. Recoveries were done at 0.5, 1.0, and 2.0 X the target concentration (1 ppm) for the recommended air volume.
Percent Recovery
Sample | 0.5x | 1.0x | 2.0x |
1 . . . . . . . . . . | 88.7 | 95.0 | 91.7 |
2 . . . . . . . . . . | 83.8 | 95.0 | 87.3 |
3 . . . . . . . . . . | 84.2 | 91.0 | 86.0 |
4 . . . . . . . . . . | 88.0 | 91.0 | 83.0 |
5 . . . . . . . . . . | 88.0 | 86.0 | 85.0 |
X . . . . . . . . . . | 86.5 | 90.5 | 87.0 |
Weighted average = 88.2
(v) Precision of the analytical procedure. The following data was used to determine the precision of the analytical method:
Concentration | 0.5 x .75 µg/mL | 1 x 1.5 µg/mL | 2 x 3.0 µg/mL |
Injection | .7421 .7441 .7831 .7753 .7612 | 1.4899 1.5826 1.4628 1.4244 1.4899 | 3.1184 3.0447 2.9149 2.9185 2.9991 |
Average Standard Deviation | .0211 | .0674 | .0998 |
CV . . . . . . . . | .0277 | .0452 | .0333 |
CV | = | 3(.0277)2 + 3 (.0452)2 + 3 (.0333)2 |
3 + 3 + 3 |
CV + 0.036 |
(vi) Storage data. Samples were generated at 1.5 mg/m3 ethylene oxide at 85% relative humidity, 22°C and 633 mm. All samples were taken for twenty minutes at 0.05 Lpm. Six samples were analyzed as soon as possible and fifteen samples were stored at refrigerated temperature (5°C) and fifteen samples were stored at ambient temperature (23°C). These stored samples were analyzed over a period of nineteen days.
Percent Recovery
Day analyzed | Refrigerated | Ambient |
1 . . . . . . . . . . . . | 87.0 | 87.0 |
1 . . . . . . . . . . . . | 93.0 | 93.0 |
1 . . . . . . . . . . . . | 94.0 | 94.0 |
1 . . . . . . . . . . . . | 92.0 | 92.0 |
4 . . . . . . . . . . . . | 92.0 | 91.0 |
4 . . . . . . . . . . . . | 93.0 | 88.0 |
4 . . . . . . . . . . . . | 91.0 | 89.0 |
6 . . . . . . . . . . . . | 92.0 | | . . . . | |
6 . . . . . . . . . . . . | 92.0 | | . . . . | |
8 . . . . . . . . . . . . | | . . . . | | 92.0 |
8 . . . . . . . . . . . . | | . . . . | | 86.0 |
10 . . . . . . . . . . . | 91.7 | | . . . . | |
10 . . . . . . . . . . . | 95.5 | | . . . . | |
10 . . . . . . . . . . . | 95.7 | | . . . . | |
11 . . . . . . . . . . . | | . . . . | | 90.0 |
11 . . . . . . . . . . . | | . . . . | | 82.0 |
13 . . . . . . . . . . . | 78.0 | | . . . . | |
13 . . . . . . . . . . . | 81.4 | | . . . . | |
13 . . . . . . . . . . . | 82.4 | | . . . . | |
14 . . . . . . . . . . . | | . . . . | | 78.5 |
14 . . . . . . . . . . . | | . . . . | | 72.1 |
18 . . . . . . . . . . . | 66.0 | | . . . . | |
18 . . . . . . . . . . . | 68.0 | | . . . . | |
19 . . . . . . . . . . . | | . . . . | | 64.0 |
19 . . . . . . . . . . . | | . . . . | | 77.0 |
(vii) Breakthrough data.
(A) Breakthrough studies were done at 2 ppm (3.6 mg/m3) at approximately 85% relative humidity at 22°C (ambient temperature). Two charcoal tubes were used in series. The backup tube was changed every ten minutes and analyzed for breakthrough. The flow rate was 0.050 Lpm.
Tube No. | Time (Minutes) | Percent breakthrough |
1 . . . . . . . . . . . . . . . . | 10 | (1) |
2 . . . . . . . . . . . . . . . . | 20 | (1) |
3 . . . . . . . . . . . . . . . . | 30 | (1) |
4 . . . . . . . . . . . . . . . . | 40 | 1.23 |
5 . . . . . . . . . . . . . . . . | 50 | 3.46 |
6 . . . . . . . . . . . . . . . . | 60 | 18.71 |
7 . . . . . . . . . . . . . . . . | 70 | 39.2 |
8 . . . . . . . . . . . . . . . . | 80 | 53.3 |
9 . . . . . . . . . . . . . . . . | 90 | 72.0 |
10 . . . . . . . . . . . . . . . | 100 | 96.0 |
11 . . . . . . . . . . . . . . . | 110 | 113.0 |
12 . . . . . . . . . . . . . . . | 120 | 133.9 |
(B) The 5% breakthrough volume was reached when 2.6 liters of test atmosphere were drawn through the charcoal tubes.
(j) References.
(i) "NIOSH Manual of Analytical Methods," 2nd ed. NIOSH: Cincinnati, 1977; Method S 286.
(ii) "IARC Monographs on the Evaluation of Carcinogenic Risk of Chemicals to Man." International Agency for Research on Cancer: Lyon, 1976; Vol. II, p. 157.
(iii) Sax., N.I. "Dangerous Properties of Industrial Materials," 4th ed.; Van Nostrand Reinhold Company, New York, 1975; p. 741.
(iv) "The Condensed Chemical Dictionary," 9th ed.; Hawley, G.G., ed.; Van Nostrand Reinhold Company, New York, 1977; p. 361.
(4) Summary of other sampling procedures. OSHA believes that several other types of monitoring equipment and techniques exist for monitoring time-weighted averages. Considerable research and method development is currently being performed, which will lead to improvements and a wider variety of monitoring techniques. A combination of monitoring procedures can be used. There probably is no one best method for monitoring personal exposure to ethylene oxide in all cases. There are advantages, disadvantages, and limitations to each method. The method of choice will depend on the need and requirements. Some commonly used methods include the use of charcoal tubes, passive dosimeters, Tedler gas sampling bags, detector tubes, photoionization detection units, infrared detection units and gas chromatographs. A number of these methods are described below.
(a) Charcoal tube sampling procedures.
(i) Qazi-Ketcham method (Ex-11-133)—This method consists of collecting EtO on Columbia JXC activated carbon, desorbing the EtO with carbon disulfide and analyzing by gas chromatography with flame ionization detection. Union Carbide has recently updated and revalidated this monitoring procedure. This method is capable of determining both eight-hour time-weighted average exposures and short-term exposures. The method was validated to 0.5 ppm. Like other charcoal collecting procedures, the method requires considerable analytical expertise.
(ii) ASTM-proposed method—The Ethylene Oxide Industry Council (EOIC) has contracted with Clayton Environmental Consultants, Inc. to conduct a collaborative study for the proposed method. The ASTM-Proposed method is similar to the method published by Qazi and Ketcham in the November 1977 American Industrial Hygiene Association Journal, and to the method of Pilney and Coyne, presented at the 1979 American Industrial Hygiene Conference. After the air to be sampled is drawn through an activated charcoal tube, the ethylene oxide is desorbed from the tube using carbon disulfide and is quantitated by gas chromatography utilizing a flame ionization detector. The ASTM-proposed method specifies a large two-section charcoal tube, shipment in dry ice, storage at less than -5°C, and analysis within three weeks to prevent migration and sample loss. Two types of charcoal tubes are being tested—Pittsburgh Coconut-Based (PCB) and columbia JXC charcoal. This collaborative study will give an indication of the inter- and intralaboratory precision and accuracy of the ASTM/proposed method. Several laboratories have considerable expertise using the Qazi-Ketcham and Dow methods.
(b) Passive monitors—Ethylene oxide diffuses into the monitor and is collected in the sampling media. The DuPont Pro-Tek badge collects EtO in an absorbing solution, which is analyzed colorimetrically to determine the amount of EtO present. The 3M 350 badge collects the EtO on chemically treated charcoal. Other passive monitors are currently being developed and tested. Both 3M and DuPont have submitted data indicating their dosimeters meet the precision and accuracy requirements of the proposed ethylene oxide standard. Both presented laboratory validation data to 0.2 ppm (Exs. 11-65, 4-20, 108, 109, 130).
(c) Tedlar gas sampling bags-samples are collected by drawing a known volume of air into a Tedlar gas sampling bag. The ethylene oxide concentration is often determined on-site using a portable gas chromatograph or portable infrared spectometer.
(d) Detector tubes—A known volume of air is drawn through a detector tube using a small hand pump. The concentration of EtO is related to the length of stain developed in the tube. Detector tubes are economical, easy to use, and give an immediate readout. Unfortunately, partly because they are nonspecific, their accuracy is often questionable. Since the sample is taken over a short period of time, they may be useful for determining the source of leaks.
(e) Direct reading instruments:
(i) There are numerous types of direct reading instruments, each having its own strengths and weaknesses (Exs. 135B, 135C, 107, 11-78, 11-153). Many are relatively new, offering greater sensitivity and specificity. Popular ethylene oxide direct reading instruments include infrared detection units, photoionization detection units, and gas chromatographs.
(ii) Portable infrared analyzers provide an immediate, continuous indication of a concentration value; making them particularly useful for locating high concentration pockets, in leak detection and in ambient air monitoring. In infrared detection units, the amount of infrared light absorbed by the gas being analyzed at selected infrared wavelengths is related to the concentration of a particular component. Various models have either fixed or variable infrared filters, differing cell pathlengths, and microcomputer controls for greater sensitivity, automation, and interference elimination.
(iii) A fairly recent detection system is photoionization detection. The molecules are ionized by high energy ultraviolet light. The resulting current is measured. Since different substances have different ionization potentials, other organic compounds may be ionized. The lower the lamp energy, the better the selectivity. As a continuous monitor, photoionization detection can be useful for locating high concentration pockets, in leak detection, and continuous ambient air monitoring. Both portable and stationary gas chromatographs are available with various types of detectors, including photoionization detectors. A gas chromatograph with a photoionization detector retains the photoionization sensitivity, but minimizes or eliminates interferences. For several GC/PID units, the sensitivity is in the 0.1-0.2 ppm EtO range. The GC/PID with microprocessors can sample up to twenty sample points sequentially, calculate and record data, and activate alarms or ventilation systems. Many are quite flexible and can be configured to meet the specific analysis needs for the workplace.
(iv) DuPont presented their laboratory validation data of the accuracy of the Qazi-Ketcham charcoal tube, the PCB charcoal tube, Miran 103 IR analyzer, 3M #3550 monitor and the DuPont C-70 badge. Quoting Elbert V. Kring:
(v) We also believe that OSHA's proposed accuracy in this standard is appropriate. At plus or minus twenty-five percent at one part per million, and plus or minus thirty-five percent below that. And, our data indicates there's only one monitoring method, right now, that we've tested thoroughly, that meets that accuracy requirements. That is the DuPont Pro-Tek badge***. We also believe that this kind of data should be confirmed by another independent laboratory, using the same type dynamic chamber testing (Tr. 1470).
Additional data by an independent laboratory following their exact protocol was not submitted. However, information was submitted on comparisons and precision and accuracy of those monitoring procedures which indicate far better precision and accuracy of those monitoring procedures than that obtained by DuPont (Ex. 4-20, 130, 11-68, 11-133, 130, 135A)
(vi) The accuracy of any method depends to a large degree upon the skills and experience of those who not only collect the samples but also those who analyze the samples. Even for methods that are collaboratively tested, some laboratories are closer to the true values than others. Some laboratories may meet the precision and accuracy requirements of the method; others may consistently far exceed them for the same method.
[Statutory Authority: Chapter
49.17 RCW. WSR 88-14-108 (Order 88-11), § 296-62-07389, filed 7/6/88; WSR 87-24-051 (Order 87-24), § 296-62-07389, filed 11/30/87.]