(1) For direct measurements performed in accordance with WAC
246-240-107, a licensee shall possess and use instrumentation to measure the activity of unsealed radioactive material before it is administered to each patient or human research subject.
(2) A licensee shall calibrate the instrumentation required in subsection (1) of this section in accordance with nationally recognized standards or the manufacturer's instructions.
(3) A licensee shall retain a record of each instrument calibration required by this section in accordance with WAC
246-240-563.
[Statutory Authority: RCW