Dose limits.
The CTDI
vol
for the following CT procedure on phantoms may not exceed the dose limits established in Table 1.
Table 1: Dose limits
CT Procedure
Phantom Size
Dose Limit:
CTDI
vol
(mGy)
Adult head
16 cm
80
Adult abdomen
32 cm
30
Pediatric head (one year old)
16 cm
40
Pediatric abdomen
(40 pounds)
16 cm
20