Design operating range (kilovolts peak) | Measured potential (kilovolts peak) | Half-value layer (milli- meters of aluminum equivalent) | Half-value layer (milli- meter of aluminum equivalent for dental units) |
Below 51–– | 30 | | 0.3 | | N/A | |
| 40 | | 0.4 | | N/A | |
| 50 | | 0.5 | | 1.5 | |
51 to 70––– | 51 | | 1.2 | | 1.5 | |
| 60 | | 1.3 | | 1.5 | |
| 70 | | 1.5 | | 1.5 | |
Above 70–– | 71 | | 2.1 | | 2.1 | |
| 80 | | 2.3 | | 2.3 | |
| 90 | | 2.5 | | 2.5 | |
| 100 | | 2.7 | | 2.7 | |
| 110 | | 3.0 | | 3.0 | |
| 120 | | 3.2 | | 3.2 | |
| 130 | | 3.5 | | 3.5 | |
| 140 | | 3.8 | | 3.8 | |
| 150 | | 4.1 | | 4.1 | |
(b) For capacitor energy storage equipment, compliance shall be determined with the system fully charged and a setting of at least 10 mAs for each exposure.
(c) The required minimal half-value layer shall include the filtration contributed by materials permanently in position between the focal spot of the tube and the patient. (For example, a table top when the tube is mounted "under the table" and inherent filtration of the tube.)
(d) Filtration control. For X-ray systems with variable kVp and variable filtration for the useful beam, a device shall link the kVp selector with the filters and shall prevent an exposure unless the minimum amount of filtration required by subdivision (a) of this subsection is in the useful beam for the selected kVp.
(6) Multiple tubes. Where two or more radiographic tubes are controlled by one exposure switch, the tube or tubes selected shall be clearly indicated prior to initiation of the exposure. Such indication shall be both on the X-ray control panel and near or on the selected tube housing assembly.
(7) Mechanical support of tube head. The tube housing assembly supports shall be adjusted such that the tube housing assembly remains stable during an exposure unless the tube housing movement during exposure is a designed function of the X-ray system.
(8) Technique indicators.
(a) The technique factors used during an exposure shall be indicated before the exposure begins, except when automatic exposure controls are used, in which case the technique factors set prior to the exposure shall be indicated.
(b) On equipment having fixed technique factors, the requirement, under subdivision (a) of this subsection may be met by permanent markings. Indication of technique factors shall be visible from the operator's position except in the case of spot films made by the fluoroscopist.
(9) Certified units. All diagnostic X-ray systems certified to comply with 21 C.F.R. 1020 shall meet the requirements of that certification.
(10) Linearity. The difference between the ratio of exposure to mAs at one mA or mAs setting and the ratio at another mA or mAs setting shall not exceed 0.10 times the sum of the ratios. This is written as:
Where X1 and X2 are the ratios (mR/mAs) for each mA or mAs station.
The test shall be performed at any selections of mA or mAs without regard to focal spot size, provided neither focal spot size is less than 0.45 millimeter.
(11) kVp accuracy. The difference between the indicated and actual kVp of an X-ray machine shall not be greater than ten percent of the indicated kVp, or, alternatively, if available, the accuracy specifications of the control panel manufacturer must be met.
[Statutory Authority: RCW
70.98.050. WSR 94-01-073, § 246-225-040, filed 12/9/93, effective 1/9/94. Statutory Authority: RCW
70.98.050 and
70.98.080. WSR 91-15-083 (Order 183), § 246-225-040, filed 7/23/91, effective 8/23/91. Statutory Authority: RCW
43.70.040. WSR 91-02-049 (Order 121), recodified as § 246-225-040, filed 12/27/90, effective 1/31/91. Statutory Authority: RCW
70.98.080. WSR 87-01-031 (Order 2450), § 402-28-035, filed 12/11/86; WSR 83-19-050 (Order 2026), § 402-28-035, filed 9/16/83. Statutory Authority: RCW
70.98.050. WSR 81-01-011 (Order 1570), § 402-28-035, filed 12/8/80; Order 1084, § 402-28-035, filed 1/14/76. Formerly WAC 402-28-030 (part).]