The CT X-ray system must:
(1) Meet the requirements of 21 C.F.R. Sec. 1020.33 (last amended June 10, 2005) at the time of installation and while the CT X-ray system is registered with the department under chapter
246-224 WAC; and
(2) Be equipped:
(a) With a visible signal that indicates when the X-ray exposure is occurring;
(b) So that the operator can immediately terminate an X-ray exposure of greater than one-half second duration at any time during the X-ray exposure;
(c) So that the parameters used during a CT procedure are:
(i) Displayed prior to beginning a scan; and
(ii) Visible by the operator from any location scanning can be initiated.
(d) So that radiation leakage from the tube port does not exceed limits established in WAC
246-225-040 (3) and (4) when data are not being collected for image production;
(e) So that the accuracy of the laser or optical positioning system is within five millimeters maximum deviation on the axial position (z-axis);
(f) With an X-ray production indicator of at least one-half second at or near the gantry that is visible from any point outside the gantry opening;
(g) So that premature termination of the X-ray exposure by the operator requires resetting of the parameters prior to initiating another scan.
[Statutory Authority: RCW
70.98.050 and
70.98.080. WSR 16-23-030, § 246-226-020, filed 11/8/16, effective 1/1/17.]