(1) Schools shall use auditory screening equipment providing tonal stimuli at frequencies at one thousand, two thousand, and four thousand herz (Hz) at hearing levels of twenty decibels (dB), as measured at the earphones, in reference to American National Standards Institute (ANSI) 1996 standards.
(2) Qualified persons will check the calibration of frequencies and intensity at least every twelve months, at the earphones, using equipment designed for audiometer calibration.